| Line 11: | Line 11: | ||
* CNR4 Net Radiometers  | * CNR4 Net Radiometers  | ||
** One for B (the original)     | ** One for B (the original)     | ||
| − | ** retap the D screws and if I can't get that to work then move the   | + | ** retap the D screws and if I can't get that to work then move the altern  | 
| + | * ate radiometer from inventory (B) to D.  | ||
| Line 104: | Line 105: | ||
* green long underwear top  | * green long underwear top  | ||
* black windproof vest  | * black windproof vest  | ||
| + | |||
| + | ===Wrap up notes===  | ||
| + | * Precip stands seem to be levelest at A & C so might be best to use them.  | ||
| + | * most net radiometers were close to level.  | ||
| + | * Look into buying more tap & die drill bits & the alternate type that dead end flat like Bill mentioned.  | ||
| + | * see about buying more of the SR50A screws that secure the disposable transducer to the main body.  | ||
| + | * The C site Net radiometer will need to be tap & drilled on the next trip.  I greased all but one screw on this trip but it was stripped on the head.  | ||
| + | * should grease all hardware (the main ones like SR50A, net radiometer etc) annually.  | ||
| + | * with things stable, should retrieve current programs and make sure git archive is up to date.  | ||
| + | |||
| + | |||
| + | ===Program note===  | ||
| + | Just a quickie.  From CR signature checker:  | ||
| + | Program Signatures Values, match with the 7th field in the first line of the .dat file  | ||
| + |      60958 GRID_SITE.CR1  | ||
| + |      27520 ngee_intensive_site_A.CR1  | ||
| + |      26235 ngee_intensive_site_B.CR1  | ||
| + |      47521 ngee_intensive_site_C.CR1  | ||
| + |       6216 ngee_intensive_site_D.CR1  | ||
| + |  majestic:barrow bob$   | ||
| + |  A matches the logger (as viewed in logger net -> status table  | ||
| + |  B matches the logger (as viewed in logger net -> status table  | ||
| + |  C matches the logger (as viewed in logger net -> status table  | ||
| + |  D matches the logger (as viewed in logger net -> status table  | ||
| + |  ... D doesn't match the data file though and I guess that is where I got hung up.  The data file didn't increment along with the change of the program signature for the D site.  | ||
| + | |||
| + | ===Post Trip Notes from Fieldbook===  | ||
| + | @ A   | ||
| + | * A trough snow mount to corroded to remove so replaced transducer in-place  | ||
| + | * Level Net Radiometer (It was close)  | ||
| + | * Level Precip (Also close to level)  | ||
| + | * Added anti-corrosion goo to net radiometer screws & connections  | ||
| + | * Changed desiccant in the box  | ||
| + | * Tools Used for Precip bucket leveling:  | ||
| + | ** 5/16" Nut Driver  | ||
| + | ** Torpedo Level  | ||
| + | * Tools Used for Snow transducer maintenance:  | ||
| + | ** 1/2" Nut Driver  | ||
| + | ** Flathead Screwdriver / 1/4" Nut Driver  | ||
| + | ** channel locks  | ||
| + | ** phillips screw driver  | ||
| + | * In-memory reset of snow depth offset  | ||
| + | ** SR50A1  = 182.1  | ||
| + | ** SR50A2 = 126.75  | ||
| + | |||
| + | @ B  | ||
| + | * Level Net Radiometer  | ||
| + | * Swap back original net radiometer  | ||
| + | * add anti-corrosion goo to radiometer & lower anemometer  | ||
| + | * level precip (it was quite a ways off)  | ||
| + | * change snow transducers  | ||
| + | * Changed desiccant in the box  | ||
| + | * In-memory reset of snow depth offset  | ||
| + | ** SR50A1  = 177.5  | ||
| + | ** SR50A2 = Unchanged from program  | ||
| + | |||
| + | @ C  | ||
| + | * C Trough snow mount to corroded to remove from the mount so transducer replaced on existing sensor  | ||
| + | * Precip still level  | ||
| + | * Net Radiation Leveled  | ||
| + | ** 1 screw is stripped on the C net radiometer so I will need to bring the tap & die up on the next trip to fix that.  | ||
| + | * Changed desiccant in the box  | ||
| + | * Anti-corrosion goo added to net radiometer and lower anemometer  | ||
| + | * In-memory reset of snow depth offset  | ||
| + | ** SR50A1  = Unchaged from program  | ||
| + | ** SR50A2 = 115.0  | ||
| + | ** SR50A3 = 117.3  | ||
| + | |||
| + | @ D  | ||
| + | * Changed net rad desiccant & leveled  | ||
| + | * swapped snow transducer, deployed S/N 2901  | ||
| + | * Think I figured out the thermal conductivity problem.  | ||
| + | * In-memory reset of snow depth offset  | ||
| + | ** SR50A1  = 159.7  | ||
| + | |||
| + | Additionally:   | ||
| + | * I think A & C precip are probably the highest quality (most stable stands).    | ||
| + | * C net radiation will need a fix for the stripped out screw next visit  | ||
| + | * Averaged 30 minutes per site.  | ||
| + | * Ended up not having wheelbarrow access but we had no problem carrying things out.  | ||
Latest revision as of 10:25, 29 August 2014
To Do:
- Change out snow transducers, 8 pieces
 - swap back the B site net radiometer
 - change out desiccant in the D site radiometer
 - level radiometers
 - Fix C site Rim Snow Sensor
 - D Thermal Conductivity
- Program works fine in Fairbanks
 - OKAY... Look at how it is wired in the field. / in pictures. I think there are two pulse port instructions when there should be just one.
 
 
To Bring:
- CNR4 Net Radiometers
- One for B (the original)
 - retap the D screws and if I can't get that to work then move the altern
 
 - ate radiometer from inventory (B) to D.
 
Plan:
Code: FDVIBG
Depart Thursday morning @ 7:40am
Return Friday night @ 12:21 AM
Contents
Thursday
- @ airport by 6ish
 - grab coke in Fairbanks airport
 - Arrive at NARL around 11am
 - get keys and dress for day in field
- wind breaker
 - backpack ready to go
 - lunch for the day
 - Water bottle, fill it
 - actionpacker with:
- 8 snow sensors
 - toolbag
 - net radiometer for B
 - drill for tapping D radiometer if needed
 - 6/32 screw for D
 - also bringing a 4/32 screw & tap for backup.
 - serial cable
 
 
 
At each site
- level radiometer
 - level precip bucket
 - look at leveling tower / anemomemters
 - listen to anemometers for bearing problems
 - swap out desiccant in the box
 - might double-check serial numbers if there is time
 - more pictures
 
Additionally
- At B swap back in original radiometer
 - At C fix the Rim snow sensor
 - At D change out the desiccant in the radiometer
 
Tools used
- Desiccant swap:
- None
 
 - Net Radiometer
- phillips screw driver
 - anti-oxidizing goo
 
 - Precip bucket
- 5/16" nut driver on hose clamp and self-tapping screw on the bucket lid
 - torpedo level
 
 - snow sensors
- Phillips screw driver (had on hand 3)
 - channel locks
 - 1/4" nut driver
 - flat screw driver
 - 1/2" nut driver
 - red electrical tape
 
 - Also had on hand:
- serial cable
 - laptop
 - voltmeter
 - cr1000 keypad
 - csi flat screw driver
 - small 1.5m tape
 - tap & die set (for net radiometer, not used this trip)
 - multitool
 - duct seal
 - cable ties
 
 
Backpack contents=
- 8 snow sensors
 - tool bag with tools above
 - first aid kit
 - rain coat
 - ultrapod
 - gopro & extra battery
 - waterproof pack cover
 - toughbook
 - extra serial cable
 - snug hat
 - mouth balaclava
 - gopro chest mount
 - this check list page
 - water bottle
 - five bar snacks
 - Extra desiccant (shoot for replacing at least annually and 2-4 packs per trip)
 - bike stereo & stereo cable plus ipod in waterproof bag
 - 2 pairs of liner gloves
 - gray fleece
 - green long underwear top
 - black windproof vest
 
Wrap up notes
- Precip stands seem to be levelest at A & C so might be best to use them.
 - most net radiometers were close to level.
 - Look into buying more tap & die drill bits & the alternate type that dead end flat like Bill mentioned.
 - see about buying more of the SR50A screws that secure the disposable transducer to the main body.
 - The C site Net radiometer will need to be tap & drilled on the next trip. I greased all but one screw on this trip but it was stripped on the head.
 - should grease all hardware (the main ones like SR50A, net radiometer etc) annually.
 - with things stable, should retrieve current programs and make sure git archive is up to date.
 
Program note
Just a quickie. From CR signature checker: Program Signatures Values, match with the 7th field in the first line of the .dat file
    60958 GRID_SITE.CR1
    27520 ngee_intensive_site_A.CR1
    26235 ngee_intensive_site_B.CR1
    47521 ngee_intensive_site_C.CR1
     6216 ngee_intensive_site_D.CR1
majestic:barrow bob$ 
A matches the logger (as viewed in logger net -> status table
B matches the logger (as viewed in logger net -> status table
C matches the logger (as viewed in logger net -> status table
D matches the logger (as viewed in logger net -> status table
... D doesn't match the data file though and I guess that is where I got hung up.  The data file didn't increment along with the change of the program signature for the D site.
Post Trip Notes from Fieldbook
@ A
- A trough snow mount to corroded to remove so replaced transducer in-place
 - Level Net Radiometer (It was close)
 - Level Precip (Also close to level)
 - Added anti-corrosion goo to net radiometer screws & connections
 - Changed desiccant in the box
 - Tools Used for Precip bucket leveling:
- 5/16" Nut Driver
 - Torpedo Level
 
 - Tools Used for Snow transducer maintenance:
- 1/2" Nut Driver
 - Flathead Screwdriver / 1/4" Nut Driver
 - channel locks
 - phillips screw driver
 
 - In-memory reset of snow depth offset
- SR50A1 = 182.1
 - SR50A2 = 126.75
 
 
@ B
- Level Net Radiometer
 - Swap back original net radiometer
 - add anti-corrosion goo to radiometer & lower anemometer
 - level precip (it was quite a ways off)
 - change snow transducers
 - Changed desiccant in the box
 - In-memory reset of snow depth offset
- SR50A1 = 177.5
 - SR50A2 = Unchanged from program
 
 
@ C
- C Trough snow mount to corroded to remove from the mount so transducer replaced on existing sensor
 - Precip still level
 - Net Radiation Leveled
- 1 screw is stripped on the C net radiometer so I will need to bring the tap & die up on the next trip to fix that.
 
 - Changed desiccant in the box
 - Anti-corrosion goo added to net radiometer and lower anemometer
 - In-memory reset of snow depth offset
- SR50A1 = Unchaged from program
 - SR50A2 = 115.0
 - SR50A3 = 117.3
 
 
@ D
- Changed net rad desiccant & leveled
 - swapped snow transducer, deployed S/N 2901
 - Think I figured out the thermal conductivity problem.
 - In-memory reset of snow depth offset
- SR50A1 = 159.7
 
 
Additionally:
- I think A & C precip are probably the highest quality (most stable stands).
 - C net radiation will need a fix for the stripped out screw next visit
 - Averaged 30 minutes per site.
 - Ended up not having wheelbarrow access but we had no problem carrying things out.