Difference between revisions of "Oliktok summer 2017 trip"

From IARC 207 Wiki
Jump to navigation Jump to search
 
(One intermediate revision by the same user not shown)
Line 10: Line 10:
 
**7)screw in all the little screws
 
**7)screw in all the little screws
 
**8)replace the sensor on the mount and screw in the cable (which is notched)
 
**8)replace the sensor on the mount and screw in the cable (which is notched)
 +
  
 
*put up temperature sensor
 
*put up temperature sensor
**mount it as close to SR50a quadpod as possible
+
**1)mount it on the south side of the fence
 +
**2)plug sensor into sensor head
 +
**3)screw
 +
**4)zip tie cable to make things neat
 +
**5)unplug logger before wiring
 
**wiring to CR1000 logger
 
**wiring to CR1000 logger
**Red    -- SW-12
+
***Red    -- SW-12
**yellow  -- SE11
+
***yellow  -- SE11
**blue    -- SE12
+
***blue    -- SE12
**clear  -- AG
+
***clear  -- AG
**black  -- AG
+
***black  -- AG
**white  -- AG
+
***white  -- AG
 
+
**replace putty
 +
**replace power to logger
  
 
*CFM100
 
*CFM100
Line 30: Line 36:
 
*level geonor
 
*level geonor
 
**there are screws at the base
 
**there are screws at the base
 +
**need to remove cover to access them- and maybe bucket as well
 
**don't do this if it doesn't seem obvious
 
**don't do this if it doesn't seem obvious
 +
 +
*visually inspect panes of LPM
 +
**IF dirty: wipe them with a paper towel or the alcohol swabbies I sent with you
  
 
*tools i'm sending you with
 
*tools i'm sending you with
Line 36: Line 46:
 
**small flathead screw driver
 
**small flathead screw driver
 
**small Philips screw driver  
 
**small Philips screw driver  
**level
+
**adjustable wrench
 +
**channel locks
  
====other===
+
===other===
 
*antifreeze
 
*antifreeze
*snip vegetation underneath
+
*<s>snip vegetation underneath
 
**cable housing
 
**cable housing
*plastic cable covers for all cables entering into the logger?
+
*plastic cable covers for all cables entering into the logger?</s>

Latest revision as of 16:14, 7 September 2017

for telayna

  • replace sr50a transducers
    • 1)unscrew sensor from mount
    • 2)unscrew cable from sensor
    • 3)unscrew all little screws holding the transducer to the sensor body (use the frisbee to work so you don't lose the little screws)
    • 4)unclip the wires of the old transducer, remove the O ring
    • 5)place desiccant in the new transducer
    • 5)place the new O ring on the sensor body and clip the new transducer wire
    • 6)line up the notch of the new transducer on the sensor body
    • 7)screw in all the little screws
    • 8)replace the sensor on the mount and screw in the cable (which is notched)


  • put up temperature sensor
    • 1)mount it on the south side of the fence
    • 2)plug sensor into sensor head
    • 3)screw
    • 4)zip tie cable to make things neat
    • 5)unplug logger before wiring
    • wiring to CR1000 logger
      • Red -- SW-12
      • yellow -- SE11
      • blue -- SE12
      • clear -- AG
      • black -- AG
      • white -- AG
    • replace putty
    • replace power to logger
  • CFM100
    • 1)unplug power from logger
    • 2)plug and screw in the CFM100
    • 3)re-plug power to the logger
    • 4)insert card
  • level geonor
    • there are screws at the base
    • need to remove cover to access them- and maybe bucket as well
    • don't do this if it doesn't seem obvious
  • visually inspect panes of LPM
    • IF dirty: wipe them with a paper towel or the alcohol swabbies I sent with you
  • tools i'm sending you with
    • 7/16 nut driver
    • small flathead screw driver
    • small Philips screw driver
    • adjustable wrench
    • channel locks

other

  • antifreeze
  • snip vegetation underneath
    • cable housing
  • plastic cable covers for all cables entering into the logger?