Difference between revisions of "Teller CAS"
Jump to navigation
Jump to search
(5 intermediate revisions by the same user not shown) | |||
Line 2: | Line 2: | ||
* N64* 58' 36.918", W166* 12' 32.67", 67 meters above sea level. | * N64* 58' 36.918", W166* 12' 32.67", 67 meters above sea level. | ||
** (Location from picture) | ** (Location from picture) | ||
+ | |||
+ | ===Data Visualizer Online=== | ||
+ | * http://monitors.iarc.uaf.edu/data/uaf_sp/teller_cas/ | ||
+ | |||
+ | ====Data Archive==== | ||
+ | * Here at some point soon (2021): | ||
+ | ** https://ngee-arctic.ornl.gov/data/#fq=datasource%3A%22NGEE%20Arctic%22&q=busey | ||
===History=== | ===History=== | ||
Line 13: | Line 20: | ||
** 2019-09-26 | ** 2019-09-26 | ||
** 2020-10-30 | ** 2020-10-30 | ||
+ | ** 2021-06-21 | ||
+ | ** 2021-09-18 | ||
===Instrumentation and Heights=== | ===Instrumentation and Heights=== | ||
Line 91: | Line 100: | ||
* 2019-09-26 23:30 | * 2019-09-26 23:30 | ||
* 2020-10-30 21:00 | * 2020-10-30 21:00 | ||
+ | * 2021-06-22 02:00 | ||
+ | * 2021-09-18 20:00 | ||
===Post Trip Notes=== | ===Post Trip Notes=== | ||
+ | ====October 2021==== | ||
+ | * In September 2021 we added a third battery and a second site to get a second snow / precip / air temperature | ||
+ | * Reviewing the data in mid October... I think the geonor is killing the battery even in the summer time. I left it connected when we left September and it seems likely to me the site will fail sooner this winter. Battery voltage had dropped to 12.2 in September after a summer of clouds. | ||
+ | |||
+ | |||
====October 2020==== | ====October 2020==== | ||
* All AT/RH Broken | * All AT/RH Broken | ||
* Think about putting out a second datalogger co-located since the wiring is so set... or not... I should look into it... | * Think about putting out a second datalogger co-located since the wiring is so set... or not... I should look into it... | ||
** ... in order to have a nonborked reference air temperature and reference tipping bucket. | ** ... in order to have a nonborked reference air temperature and reference tipping bucket. | ||
− | + | * Reviewing pictures 10/2021 while trying to figure out when the radiometer went out of level I see it was sometime before the October 2020 data download. | |
====June 2019==== | ====June 2019==== | ||
Line 113: | Line 129: | ||
===Data Notes=== | ===Data Notes=== | ||
− | Most Recent QA : 2021- | + | Most Recent QA : 2021-10-21 |
====Barometric Pressure==== | ====Barometric Pressure==== | ||
− | * | + | * Mostly Good |
* NAN data is due to low battery voltage | * NAN data is due to low battery voltage | ||
+ | * I don't know why but replacing the broken AT/RH and replacing them with working ones bonked this sensor for the summer. I didn't notice till 10/21 but can see from the data that | ||
====1.5 meter Air Temperature / Relative Humidity / Vapor Pressure==== | ====1.5 meter Air Temperature / Relative Humidity / Vapor Pressure==== | ||
* Sensor Fails over 2018-2019 winter due to moisture. | * Sensor Fails over 2018-2019 winter due to moisture. | ||
Line 134: | Line 151: | ||
====Precipitation==== | ====Precipitation==== | ||
* Disconnected in winter to reduce battery load. | * Disconnected in winter to reduce battery load. | ||
+ | * Need to empty bucket every visit. It looks like there is just too much precip in Nome area for our visiting frequency. | ||
====Radiation==== | ====Radiation==== |
Latest revision as of 14:25, 20 October 2021
Contents
Location
- N64* 58' 36.918", W166* 12' 32.67", 67 meters above sea level.
- (Location from picture)
Data Visualizer Online
Data Archive
- Here at some point soon (2021):
History
- Site installation 9/21 and 9/24 2018.
- Final initial visit on 9/26/18
- Pictures are filed in "2018 Work Nome September 2018" including pictures of fieldbook notes.
- Additional Site visits:
- 2019-06-10
- 2019-06-16
- 2019-09-19
- 2019-09-26
- 2020-10-30
- 2021-06-21
- 2021-09-18
Instrumentation and Heights
- Barometric Air Pressure
- Vaisala PTB110
- https://www.campbellsci.com/cs106
- Range: 500 to 1100 hPa
- Accuracy: +-1.5 hPa -40 to +60 C
- AT/RH @ 1.5m, 3.0m
- Rotronic HygroClip2 HC2A
- https://www.campbellsci.com/hc2s3
- -40 to +60 Celsius
- =-0.1 C at 23 C
- 0 to 100 % Relative Humidity
- +-0.8% at 23C
- WS/WD @ 1.5m, 3.0m
- RM Young 01508
- https://www.campbellsci.com/05108-45-l
- Range: 0 to 100 m/s
- Accuracy: +-0.3m/s
- Startup threshold: 1.0 m/s
- Wind Direction measurement band: 0 to 355 degrees
- Accuracy: +-3 degrees
- Snow Depth @ 3.0
- Campbell Scientific SR50A
- https://www.campbellsci.com/sr50a
- Resolution: 0.25mm
- Accuracy: +-1.2 cm (0.4% of 3.0 meter measurement range)
- Net Radiation @ 1.5m
- Kipp & Zonen CNR4
- https://www.campbellsci.com/cnr4
- Shortwave Measurement Range- 305 to 2800 nm
- Shortwave Uncertainty in total daily - < 5% (The uncertainty values are for a 95% confidence level.)
- Longwave Measurement Range - 4500 to 42,000 nm
- Longwave Uncertainty in total daily- < 10% (The uncertainty values are for a 95% confidence level.)
- Soil Pit 1 = Dry
- Campbell Scientific CS109 sensors
- https://www.campbellsci.com/109
- Range: -50° to +70°C
- Tolerance: ±0.2°C (over 0° to 70°C range)
- Interchangeability Error: ±0.1°C (over 0° to 70°C range increasing to ±0.5°C at -50°C)
- 1.5cm bgs
- 10cm bgs
- 20cm bgs
- 40cm bgs
- 80cm bgs
- Campbell Scientific CS109 sensors
- Soil Pit 2 = Wet
- Campbell Scientific CS109 sensors
- https://www.campbellsci.com/109
- Range: -50° to +70°C
- Tolerance: ±0.2°C (over 0° to 70°C range)
- Interchangeability Error: ±0.1°C (over 0° to 70°C range increasing to ±0.5°C at -50°C)
- 5cm bgs
- 10cm bgs
- 40cm bgs
- 80cm bgs
- Campbell Scientific CS109 sensors
- All season Precipitation:
- Geonor Weighing Bucket
- https://www.geonor.com/brochures/t-200b-series-all-weather.pdf
- Collecting Area - 200 cm^2
- Capacity - 1500 mm H2O (including antifreeze)
- Sensitivity - 0.05mm
- Accuracy - 0.1% FS
- ... only powered for summer season
- Drained in October 2020
- Skin Temperature @ 1.5m:
- Apogee Instruments SI-111
- https://www.campbellsci.com/si-111
- Operating Range: -55 to +80 C
- Accuracy: ±0.2°C (-10° to +65°C) & ±0.5°C (-40° to +70°C)
- Field of View: 22° half angle
Data Downloads
- Station turned on, first record:
- 2018-09-24 22:00
- 2019-06-10 21:30
- 2019-09-26 23:30
- 2020-10-30 21:00
- 2021-06-22 02:00
- 2021-09-18 20:00
Post Trip Notes
October 2021
- In September 2021 we added a third battery and a second site to get a second snow / precip / air temperature
- Reviewing the data in mid October... I think the geonor is killing the battery even in the summer time. I left it connected when we left September and it seems likely to me the site will fail sooner this winter. Battery voltage had dropped to 12.2 in September after a summer of clouds.
October 2020
- All AT/RH Broken
- Think about putting out a second datalogger co-located since the wiring is so set... or not... I should look into it...
- ... in order to have a nonborked reference air temperature and reference tipping bucket.
- Reviewing pictures 10/2021 while trying to figure out when the radiometer went out of level I see it was sometime before the October 2020 data download.
June 2019
- Tripod Mast Bent
- CNR4 difficult to level
- Could use a some help keeping lower anemometer from moving. Like maybe a couple pieces of unistrut to keep arm from wobbling
- Need to replace lower AT/RH as it seems completely broken.
- Rotronic HygroClip2 HC2A
- Should look to see if they sell a better top cap for it more resistant to moisture
- Need to get some SW12V adapter to turn AT/RH on/off and look at program / wiring for other things that could be turned off.
- Left 2 guy wire anchors and 100' of guy wire at UAF Storage. Could bring out more anchors or unistrut / nails or something. Maybe pound unistrut into ground and then pin in place?
- I thought Geonor and Solar Panel would require guying but both survived the winter while tripod mast didn't.
- I think snow sensor maybe failed over the winter, too.
- I was thinking a heavier gauge cable for solar panel might be appropriate, too.
Data Notes
Most Recent QA : 2021-10-21
Barometric Pressure
- Mostly Good
- NAN data is due to low battery voltage
- I don't know why but replacing the broken AT/RH and replacing them with working ones bonked this sensor for the summer. I didn't notice till 10/21 but can see from the data that
1.5 meter Air Temperature / Relative Humidity / Vapor Pressure
- Sensor Fails over 2018-2019 winter due to moisture.
- 2019-01-03 23:00 UTC
3.0 meter Air Temperature / Relative Humidity / Vapor Pressure
- NAN data in 2019 Winter is due to low battery voltage
- Without site visit hard to say why sensor goes dead on 2019-09-19 21:00
- Loose wire?
- the timing...
1.5 meter Anemometer
- NAN data in Winter is due to low battery voltage
3.0 meter Anemometer
- NAN data in Winter is due to low battery voltage
Precipitation
- Disconnected in winter to reduce battery load.
- Need to empty bucket every visit. It looks like there is just too much precip in Nome area for our visiting frequency.
Radiation
- Albedo
- Blip at 2018-11-20 3:30 to delete
- need to limit to 0. to 1.
Soil Pit 1
- Winter data gaps and NANs due to low battery voltage
- Sensor 5 dies 2020-05-16 22:00
- Sensor 3 dies 2020-05-16 22:00
- Sensor 2 dies 2020-06-01 10:30
- Sensor 4 dies 2020-06-01 10:30
Soil Pit 2
- Winter data gaps and NANs due to low battery voltage
Snow
- Transducer fails in January 2019.