Difference between revisions of "Spring 2016 site visit"

From IARC 207 Wiki
Jump to navigation Jump to search
Line 1: Line 1:
*take photos of sensors without pics on instrumentation site
+
 
 
*bring measuring cup set for leveling UVA and UVB sensors
 
*bring measuring cup set for leveling UVA and UVB sensors
 
*wire junction to split all instruments needing 12V from logger
 
*wire junction to split all instruments needing 12V from logger
Line 51: Line 51:
 
** to turn on:press and hold organge power button until green LED lights up. after a few sec beep twice.  green LED every 15 sec indicates normal running. if any other beeping or lights, consult manual
 
** to turn on:press and hold organge power button until green LED lights up. after a few sec beep twice.  green LED every 15 sec indicates normal running. if any other beeping or lights, consult manual
  
 
+
== take photos of sensors without pics on instrumentation site==
==To change on instrumentation website==
+
*HMP155
** pluvio is actually version 2
+
*wind sensors
** lake temp measurement
+
*snow dept
** barometer
+
*barometer
 +
*evap pan
  
 
==program changes==
 
==program changes==
Line 62: Line 63:
 
*possible pluvio calibration
 
*possible pluvio calibration
 
*UVA+UVB corrections
 
*UVA+UVB corrections
 +
 +
==To change on instrumentation website==
 +
** pluvio is actually version 2
 +
** lake temp measurement
 +
** barometer
  
  

Revision as of 10:49, 3 June 2016

  • bring measuring cup set for leveling UVA and UVB sensors
  • wire junction to split all instruments needing 12V from logger
  • swap in fancier wind sensor
  • fill in serial numbers of all instruments/loggers on google sheet
  • swap out HMP155s
    • write down which height is which SN
  • swap out quantum sensors
    • bring mini wrench to replace them
    • figure out what the connection is like at toolik
    • change calibration constant?
  • swap CNR4
  • remove old radio
  • get field books from Brie
  • install new barometer
    • change program calibration constants
  • change UVA + UVB program calibration constants: UVA:0.0221, UVB: 0.00178 (per Lija's email)
  • we zero out the CMP6, but not the CNR4. reconsider?

install new evap pan

    • leak test (marine epoxy in case of leaks. should be up in the EDC office)
    • clean out bottom
    • field calibrate

Calibrate

   Fill pan partway with water, measure height with ruler in mm
   Measure voltage by reading the EvapLevelTemp variable on the program
   Add more water, measure water depth in pan
   Measure voltage
   Calculate multiplier and offset using y = mx+b
       m(multiplier) = (height1-height2)/(voltage2-voltage1)
       b(offset) = height1-(m*voltage1) 

Change multiplier and offset in program


calibrate precip

Tipping bucket

    • fill bottle to line, use smallest orifice, do this 3 times
    • should be 30mm per bottle, so 90mm total
    • change multiplier in program if necessary

Pluvio accuracy test

    • received software in ftp from ott (on my fancy laptop, need to transfer to toughbook)
    • need usb cable

download garden cam images

    • bring replacement batteries
    • to turn off: press and hold organge power button until 3 beeps and green LED turns off
    • remove drive, download and clear images on computer
    • to turn on:press and hold organge power button until green LED lights up. after a few sec beep twice. green LED every 15 sec indicates normal running. if any other beeping or lights, consult manual

take photos of sensors without pics on instrumentation site

  • HMP155
  • wind sensors
  • snow dept
  • barometer
  • evap pan

program changes

  • evap pan calibration
  • possible tipping bucket calibration
  • possible pluvio calibration
  • UVA+UVB corrections

To change on instrumentation website

    • pluvio is actually version 2
    • lake temp measurement
    • barometer


to bring

    • laptop + charger
    • field laptop + charger
    • tipping bucket calibration box
    • gloves, hat(warm and sun)
    • tools
    • wire junction
    • hiking boots,rubber boots,running shoes
    • field book
    • usb cable for pluvio
    • batteries for garden cam
    • cookies for brie
    • instruments
      • anemometer
      • CNR4
      • Two HMP155s