Location

  • N64* 58' 36.918", W166* 12' 32.67", 67 meters above sea level.
    • (Location from picture)

Data Visualizer Online

Data Archive

History

  • Site installation 9/21 and 9/24 2018.
  • Final initial visit on 9/26/18
  • Pictures are filed in "2018 Work Nome September 2018" including pictures of fieldbook notes.
  • Additional Site visits:
    • 2019-06-10
    • 2019-06-16
    • 2019-09-19
    • 2019-09-26
    • 2020-10-30
    • 2021-06-21
    • 2021-09-18

Instrumentation and Heights

  • Barometric Air Pressure
  • AT/RH @ 1.5m, 3.0m
  • WS/WD @ 1.5m, 3.0m
    • RM Young 01508
    • https://www.campbellsci.com/05108-45-l
    • Range: 0 to 100 m/s
    • Accuracy: +-0.3m/s
    • Startup threshold: 1.0 m/s
    • Wind Direction measurement band: 0 to 355 degrees
    • Accuracy: +-3 degrees
  • Snow Depth @ 3.0
  • Net Radiation @ 1.5m
    • Kipp & Zonen CNR4
    • https://www.campbellsci.com/cnr4
    • Shortwave Measurement Range- 305 to 2800 nm
    • Shortwave Uncertainty in total daily - < 5% (The uncertainty values are for a 95% confidence level.)
    • Longwave Measurement Range - 4500 to 42,000 nm
    • Longwave Uncertainty in total daily- < 10% (The uncertainty values are for a 95% confidence level.)
  • Soil Pit 1 = Dry
    • Campbell Scientific CS109 sensors
      • https://www.campbellsci.com/109
      • Range: -50° to +70°C
      • Tolerance: ±0.2°C (over 0° to 70°C range)
      • Interchangeability Error: ±0.1°C (over 0° to 70°C range increasing to ±0.5°C at -50°C)
    • 1.5cm bgs
    • 10cm bgs
    • 20cm bgs
    • 40cm bgs
    • 80cm bgs
  • Soil Pit 2 = Wet
    • Campbell Scientific CS109 sensors
      • https://www.campbellsci.com/109
      • Range: -50° to +70°C
      • Tolerance: ±0.2°C (over 0° to 70°C range)
      • Interchangeability Error: ±0.1°C (over 0° to 70°C range increasing to ±0.5°C at -50°C)
    • 5cm bgs
    • 10cm bgs
    • 40cm bgs
    • 80cm bgs
  • All season Precipitation:
  • Skin Temperature @ 1.5m:
    • Apogee Instruments SI-111
    • https://www.campbellsci.com/si-111
    • Operating Range: -55 to +80 C
    • Accuracy: ±0.2°C (-10° to +65°C) & ±0.5°C (-40° to +70°C)
    • Field of View: 22° half angle

Data Downloads

  • Station turned on, first record:
    • 2018-09-24 22:00
  • 2019-06-10 21:30
  • 2019-09-26 23:30
  • 2020-10-30 21:00
  • 2021-06-22 02:00
  • 2021-09-18 20:00

Post Trip Notes

October 2020

  • All AT/RH Broken
  • Think about putting out a second datalogger co-located since the wiring is so set... or not... I should look into it...
    • ... in order to have a nonborked reference air temperature and reference tipping bucket.


June 2019

  • Tripod Mast Bent
  • CNR4 difficult to level
  • Could use a some help keeping lower anemometer from moving. Like maybe a couple pieces of unistrut to keep arm from wobbling
  • Need to replace lower AT/RH as it seems completely broken.
    • Rotronic HygroClip2 HC2A
    • Should look to see if they sell a better top cap for it more resistant to moisture
  • Need to get some SW12V adapter to turn AT/RH on/off and look at program / wiring for other things that could be turned off.
  • Left 2 guy wire anchors and 100' of guy wire at UAF Storage. Could bring out more anchors or unistrut / nails or something. Maybe pound unistrut into ground and then pin in place?
  • I thought Geonor and Solar Panel would require guying but both survived the winter while tripod mast didn't.
  • I think snow sensor maybe failed over the winter, too.
  • I was thinking a heavier gauge cable for solar panel might be appropriate, too.

Data Notes

Most Recent QA : 2021-02-11

Barometric Pressure

  • All Good
  • NAN data is due to low battery voltage

1.5 meter Air Temperature / Relative Humidity / Vapor Pressure

  • Sensor Fails over 2018-2019 winter due to moisture.
  • 2019-01-03 23:00 UTC

3.0 meter Air Temperature / Relative Humidity / Vapor Pressure

  • NAN data in 2019 Winter is due to low battery voltage
  • Without site visit hard to say why sensor goes dead on 2019-09-19 21:00
    • Loose wire?
    • the timing...

1.5 meter Anemometer

  • NAN data in Winter is due to low battery voltage

3.0 meter Anemometer

  • NAN data in Winter is due to low battery voltage

Precipitation

  • Disconnected in winter to reduce battery load.

Radiation

  • Albedo
    • Blip at 2018-11-20 3:30 to delete
    • need to limit to 0. to 1.

Soil Pit 1

  • Winter data gaps and NANs due to low battery voltage
  • Sensor 5 dies 2020-05-16 22:00
  • Sensor 3 dies 2020-05-16 22:00
  • Sensor 2 dies 2020-06-01 10:30
  • Sensor 4 dies 2020-06-01 10:30

Soil Pit 2

  • Winter data gaps and NANs due to low battery voltage

Snow

  • Transducer fails in January 2019.