Oliktok summer 2017 trip
Revision as of 13:52, 7 September 2017 by 137.229.92.185 (talk)
for telayna
- replace sr50a transducers
- 1)unscrew sensor from mount
- 2)unscrew cable from sensor
- 3)unscrew all little screws holding the transducer to the sensor body (use the frisbee to work so you don't lose the little screws)
- 4)unclip the wires of the old transducer, remove the O ring
- 5)place desiccant in the new transducer
- 5)place the new O ring on the sensor body and clip the new transducer wire
- 6)line up the notch of the new transducer on the sensor body
- 7)screw in all the little screws
- 8)replace the sensor on the mount and screw in the cable (which is notched)
- put up temperature sensor
- mount it as close to SR50a quadpod as possible
- wiring to CR1000 logger
- Red -- SW-12
- yellow -- SE11
- blue -- SE12
- clear -- AG
- black -- AG
- white -- AG
- CFM100
- 1)unplug power from logger
- 2)plug and screw in the CFM100
- 3)re-plug power to the logger
- 4)insert card
- level geonor
- there are screws at the base
- don't do this if it doesn't seem obvious
- tools i'm sending you with
- 7/16 nut driver
- small flathead screw driver
- small Philips screw driver
- level
=other
- antifreeze
- snip vegetation underneath
- cable housing
- plastic cable covers for all cables entering into the logger?