Teller CAS
Contents
Location
- N64* 58' 36.918", W166* 12' 32.67", 67 meters above sea level.
- (Location from picture)
Data Visualizer Online
Data Archive
- Here at some point soon (2021):
History
- Site installation 9/21 and 9/24 2018.
- Final initial visit on 9/26/18
- Pictures are filed in "2018 Work Nome September 2018" including pictures of fieldbook notes.
- Additional Site visits:
- 2019-06-10
- 2019-06-16
- 2019-09-19
- 2019-09-26
- 2020-10-30
Instrumentation and Heights
- Barometric Air Pressure
- Vaisala PTB110
- https://www.campbellsci.com/cs106
- Range: 500 to 1100 hPa
- Accuracy: +-1.5 hPa -40 to +60 C
- AT/RH @ 1.5m, 3.0m
- Rotronic HygroClip2 HC2A
- https://www.campbellsci.com/hc2s3
- -40 to +60 Celsius
- =-0.1 C at 23 C
- 0 to 100 % Relative Humidity
- +-0.8% at 23C
- WS/WD @ 1.5m, 3.0m
- RM Young 01508
- https://www.campbellsci.com/05108-45-l
- Range: 0 to 100 m/s
- Accuracy: +-0.3m/s
- Startup threshold: 1.0 m/s
- Wind Direction measurement band: 0 to 355 degrees
- Accuracy: +-3 degrees
- Snow Depth @ 3.0
- Campbell Scientific SR50A
- https://www.campbellsci.com/sr50a
- Resolution: 0.25mm
- Accuracy: +-1.2 cm (0.4% of 3.0 meter measurement range)
- Net Radiation @ 1.5m
- Kipp & Zonen CNR4
- https://www.campbellsci.com/cnr4
- Shortwave Measurement Range- 305 to 2800 nm
- Shortwave Uncertainty in total daily - < 5% (The uncertainty values are for a 95% confidence level.)
- Longwave Measurement Range - 4500 to 42,000 nm
- Longwave Uncertainty in total daily- < 10% (The uncertainty values are for a 95% confidence level.)
- Soil Pit 1 = Dry
- Campbell Scientific CS109 sensors
- https://www.campbellsci.com/109
- Range: -50° to +70°C
- Tolerance: ±0.2°C (over 0° to 70°C range)
- Interchangeability Error: ±0.1°C (over 0° to 70°C range increasing to ±0.5°C at -50°C)
- 1.5cm bgs
- 10cm bgs
- 20cm bgs
- 40cm bgs
- 80cm bgs
- Campbell Scientific CS109 sensors
- Soil Pit 2 = Wet
- Campbell Scientific CS109 sensors
- https://www.campbellsci.com/109
- Range: -50° to +70°C
- Tolerance: ±0.2°C (over 0° to 70°C range)
- Interchangeability Error: ±0.1°C (over 0° to 70°C range increasing to ±0.5°C at -50°C)
- 5cm bgs
- 10cm bgs
- 40cm bgs
- 80cm bgs
- Campbell Scientific CS109 sensors
- All season Precipitation:
- Geonor Weighing Bucket
- https://www.geonor.com/brochures/t-200b-series-all-weather.pdf
- Collecting Area - 200 cm^2
- Capacity - 1500 mm H2O (including antifreeze)
- Sensitivity - 0.05mm
- Accuracy - 0.1% FS
- ... only powered for summer season
- Drained in October 2020
- Skin Temperature @ 1.5m:
- Apogee Instruments SI-111
- https://www.campbellsci.com/si-111
- Operating Range: -55 to +80 C
- Accuracy: ±0.2°C (-10° to +65°C) & ±0.5°C (-40° to +70°C)
- Field of View: 22° half angle
Data Downloads
- Station turned on, first record:
- 2018-09-24 22:00
- 2019-06-10 21:30
- 2019-09-26 23:30
- 2020-10-30 21:00
Post Trip Notes
October 2020
- All AT/RH Broken
- Think about putting out a second datalogger co-located since the wiring is so set... or not... I should look into it...
- ... in order to have a nonborked reference air temperature and reference tipping bucket.
June 2019
- Tripod Mast Bent
- CNR4 difficult to level
- Could use a some help keeping lower anemometer from moving. Like maybe a couple pieces of unistrut to keep arm from wobbling
- Need to replace lower AT/RH as it seems completely broken.
- Rotronic HygroClip2 HC2A
- Should look to see if they sell a better top cap for it more resistant to moisture
- Need to get some SW12V adapter to turn AT/RH on/off and look at program / wiring for other things that could be turned off.
- Left 2 guy wire anchors and 100' of guy wire at UAF Storage. Could bring out more anchors or unistrut / nails or something. Maybe pound unistrut into ground and then pin in place?
- I thought Geonor and Solar Panel would require guying but both survived the winter while tripod mast didn't.
- I think snow sensor maybe failed over the winter, too.
- I was thinking a heavier gauge cable for solar panel might be appropriate, too.
Data Notes
Most Recent QA : 2021-02-11
Barometric Pressure
- All Good
- NAN data is due to low battery voltage
1.5 meter Air Temperature / Relative Humidity / Vapor Pressure
- Sensor Fails over 2018-2019 winter due to moisture.
- 2019-01-03 23:00 UTC
3.0 meter Air Temperature / Relative Humidity / Vapor Pressure
- NAN data in 2019 Winter is due to low battery voltage
- Without site visit hard to say why sensor goes dead on 2019-09-19 21:00
- Loose wire?
- the timing...
1.5 meter Anemometer
- NAN data in Winter is due to low battery voltage
3.0 meter Anemometer
- NAN data in Winter is due to low battery voltage
Precipitation
- Disconnected in winter to reduce battery load.
Radiation
- Albedo
- Blip at 2018-11-20 3:30 to delete
- need to limit to 0. to 1.
Soil Pit 1
- Winter data gaps and NANs due to low battery voltage
- Sensor 5 dies 2020-05-16 22:00
- Sensor 3 dies 2020-05-16 22:00
- Sensor 2 dies 2020-06-01 10:30
- Sensor 4 dies 2020-06-01 10:30
Soil Pit 2
- Winter data gaps and NANs due to low battery voltage
Snow
- Transducer fails in January 2019.